5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1136

■■■■■■■■

tiamo 2.4

from other measuring command) by which every measured absorbance
value can then be corrected.

Input range

–4,000 - 4,000 mAU

Default value

0 mAU

Integration time

Integration time for absorbance measurement.

Input range

0.01 - 600,000 ms

Default value

6 ms

Averaged spectra

Number of spectra that are recorded and averaged.

Input range

1 - 10,000

Default value

10

Smoothing

Number of neighboring pixels with which the light intensity of each mea-
surement pixel is determined.

Input range

1 - 100 pixels

Default value

1 pixels

Flash rate

Number of flashes per second of a pulsed xenon lamp as light source.

Input range

0 - 100 Hz

Selection

off

Default value

off

off
For the continuous type of the light source.

Measuring parameters for the 089 Photometer

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.

Input range

0 - 4,000 mAU

Default value

0 mAU

Measurement
Two types of measurement can be selected, Measurement with drift
control
or Measurement without drift control.