5 Method

tiamo 2.4

■■■■■■■■

837

Selection

maximum

Min. rate

This parameter defines the minimum rate with which iodine is generated
inside the Control range. The smaller the rate the slower the titration
and the higher the accuracy.

756, 831
Input range

0.3 - 999.9 µg/min

Default value

0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min

Selection

minimum

851, 852
Input range

0.3 - 999.0 µg/min

Default value

0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min

Selection

minimum

Stop criterion

Stop criterion

Switches off titration when the endpoint has been reached and the stop
criterion has been met.

756, 831
Selection

Drift | rel. drift

Default value

rel. drift

851, 852
Selection

Drift | rel. drift | Time | off

Default value

rel. drift

Stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.

Input range

1 - 999 µg/min

Default value

5 µg/min

Relative stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion
= rel. drift.

Input range

0 - 999 µg/min

Default value

5 µg/min