Metrohm tiamo 2.4 Manual User Manual

Page 817

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

801

Titrando
Input range

1 - 999 µg/min

Default value

20 µg/min

Drift correction

If this parameter is set to automatic or manual the drift correction ( =
Drift value × Time for drift correction) caused by empty consumption
during titration is deducted from the total consumption. The Time for
drift correction
is the period between end of conditioning and end of
titration.

Selection

automatic | manual | off

Default value

off

automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Input range

0.0 - 99.9 µL/min

Default value

0.0 µL/min

Stop volume

Stop of the determination when the specified volume has been dosed
since the start of the conditioning. The stop volume should be adjusted to
the size of the titration cell in order to prevent any overflow.

Titrando
Input range

0.000000 - 9,999.99 mL

Default value

20.000 mL

Selection

off

Default value

off

Stop time

Stop of the determination when the specified time has elapsed since the
start of the conditioning.

Titrando
Input range

1 - 999,999 s