5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

706

■■■■■■■■

tiamo 2.4

Control

NOTE

Will be displayed only if the Users option is selected in the selection list
Titration rate.

Dynamics pH

The control range defines the measured value range before the specified
endpoint. It has a decisive influence on the titration rate and thus also on
the accuracy. Within the control range, dosing is controlled by the Min.
rate
. The closer the endpoint, the slower the dosing until the Min. rate
has been reached. The larger the control range, the slower the titration.
Outside the control range, dosing is carried out continuously at Max.
rate
(see Chapter 5.6.3.4.2, page 695).

Titrando, 855
Input range

0.001 - 20.000

Default value

5.000 (slow), 2.000 (optimal), 0.500 (fast)

Selection

off

off
The entire measured value range is defined as control range. The mini-
mum titration rate is applied throughout the entire titration.

Titrino
Input range

0.01 - 20.00

Default value

5.00 (slow), 2.00 (optimal), 0.50 (fast)

Selection

off

off
The entire measured value range is defined as control range. The mini-
mum titration rate is applied throughout the entire titration.

Max. rate

Rate at which dosing is carried out outside of the control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Default value

1.00 (slow), 10.00 (optimal), 166.00
(fast) mL/min

Selection

maximum