5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

548

■■■■■■■■

tiamo 2.4

Input range

0 - 999999 s

Default value

1 s

Start volume

Start volume

Volume to be added before the start of the titration at the dosing rate
indicated.

Titrando, 855
Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Titrino
Input range

0.00 - 999.99 mL

Default value

0.00 mL

Dosing rate

Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

Start measured value

NOTE

Only displayed for Titrando and 855.

Start measured value

Once the start measured value is reached, the dosing of the start volume
is stopped and the next start condition is processed or the titration is
started. If the start measured value is achieved by the addition of a start
volume, then titration will start directly.

Input range

-200.0 - 200.0 µA