5 Method

tiamo 2.4

■■■■■■■■

1155

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Device types | Titrando

Default value

Titrando

Sensor

Measuring input

Selection of the measuring input the sensor is connected to.

Titrando, 867
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Titrino
Selection

1 | 2 | diff

Default value

1

Sensor

Selection of a sensor of type pH electrode from the list of sensors avail-
able in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.

Selection

Sensor name | pH electrode

Default value

pH electrode

Temperature measurement

Type of temperature measurement.

Titrando, 855, 867
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.