Metrohm tiamo 2.4 Manual User Manual

Page 771

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

755

The conditions for stopping the titration are defined on this tab. If several
stop conditions are set, then the criterion which is fulfilled first will stop
the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Titrando, 855
Input range

0.00000 - 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Input range

0.000 - 9999.99 mL

Default value

100.00 mL

Selection

off

Stop time

Stops when the preset time (including start conditions) has elapsed since
the start of the titration.

Titrando, 855
Input range

0 - 999999 s

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum