5 Method

tiamo 2.4

■■■■■■■■

1141

If these buffers are used, tiamo will recognize the specific buffer automati-
cally.

Selection

Metrohm | NIST | DIN | Fisher | Fluka Basel |
Mettler | Merck CertiPUR 20°C/Titrisol |
Merck CertiPUR 25°C | Beckman | Custom |
Special | Radiometer | Baker | Hamilton
DURACAL
| Precisa

Default value

Metrohm

Custom
The custom buffer series defined in the configuration under
Tools

Templates Custom calibration buffers is used.

Special
Up to nine calibration buffers can be defined directly in the properties
window of the CAL LOOP pH command.

NOTE

The automatic buffer recognition is deactivated for the buffer type Spe-
cial
, and the buffers must therefore be measured at the time of calibra-
tion in the same order as they were defined in the CAL LOOP pH com-
mand.

Buffer # pH

pH value of the special buffers (displayed only for Buffer type = Special).

Input range

–20.000 - 20.000

NOTE

Enter the pH values for the calibration temperature used. A message
will be displayed correspondingly often during the run to indicate the
next buffer to be measured.

Request for buffer exchange

on | off (Default value: on)

If this check box is activated, a request for manual buffer exchange is dis-
played after each time the calibration loop is run through. This parameter
must be disabled for automatic buffer exchange when calibrating with a
sample changer.