5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1040
■■■■■■■■
tiamo 2.4
Integration time
Integration time for spectrum recording.
Input range
0.01 - 600000 ms
Default value
6 ms
Averaged spectra
Number of spectra that are recorded and averaged.
Input range
1 - 10000
Default value
10
Smoothing
Number of neighboring pixels with which the value for each measurement
pixel is determined.
Input range
0 - 100 Pixels
Default value
0 Pixels
Flash rate
Number of flashes per second of a pulsed xenon lamp as light source.
Input range
0 - 100 Hz
Selection
off
Default value
off
off
For the continuous type of the light source.
5.6.4.13
MEAS Spec
5.6.4.13.1
MEAS Spec - Overview
Dialog window: Method
▶ MEAS Spec ▶ Properties... ▶ MEAS Spec - 'Command
name'
Command for the measurement of the sample solution across a
defined wavelength range.
Devices
This command can be executed with the following instrument:
Spectrometer: Avantes