Metrohm tiamo 2.4 Manual User Manual

Page 1115

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1099

* Definition of the number of calibration solutions and the type of solution
exchange.

** Move to the next calibration solution.
 

The request for changing the buffer or the standard must be deactivated
in the command CAL LOOP pH or CAL LOOP Conc, respectively. No
request for changing the buffer or standard is then displayed during the
run.

For the automatic move to the next calibration solution on the sample
changer rack, a MOVE command must be inserted between the loop
command CAL LOOP and the measuring command CAL MEAS. The
positions of the calibration solutions on the rack are defined in this com-
mand. The following three possibilities are available to do so: