Metrohm tiamo 2.4 Manual User Manual

Page 926

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

910

■■■■■■■■

tiamo 2.4

Upper hysteresis

Upper hysteresis of the dosing rate. If the dosing rate once again falls
below the upper limit by this hysteresis value, then the event Dosing rate
upper limit OK
is triggered.

Titrando, 855
Input range

0.00 - 166.00 mL/min

Default value

0.20 mL/min

Action

Selection of the event to be triggered if the upper or lower limit is viola-
ted:

Selection

Cancel determination | Cancel command
(Titrando, 855)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT U command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando, 855)
The running STAT U command will be canceled, then the next com-
mand will be carried out.
Wait for [Continue]
Reagent dosing in the running STAT U command will be interrupted
and a message will be displayed. As soon as the monitored measured
value is again within the limits (including hysteresis), reagent dosing
can be resumed by pressing [Continue] in this message box.
Wait for limit ok
Reagent dosing in the running STAT U command will be interrupted.
As soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are violated.

Monitoring Temperature

Monitoring Temperature

on | off (Default value: off)

If this option is enabled, then the temperature will be monitored and any
violated limits will be entered in the measuring point list.

Advertising