2 options - general – Metrohm tiamo 2.4 Manual User Manual

Page 1366

Advertising
background image

6.3 Configuration data

■■■■■■■■■■■■■■■■■■■■■■

1350

■■■■■■■■

tiamo 2.4

6.3.4.2

Options - General

Tab: Configuration

Tools Options... Options General

Dialog language

Dialog language

Selection of the dialog language.

Selection

German | English | Additional languages
(dependent on installed language patches)

Default value

English

NOTE

The program must be restarted for the modified setting to become
effective.

Emergency stop button
Options for the display of the Emergency stop tiamo button.

Selection

On | Off

Default value

Off

On
The button is displayed in all program parts.
Off
The button is not displayed.

The button can be moved to any position with the left mouse button
pressed down and is always shown at the very top of the screen both in
the program window and outside it on the Windows desktop. A mouse
click on this button immediately stops all running determinations on all
active workplaces. When this happens, all devices are stopped (including
pumps; any device commands that may have been started will be carried
out to completion, e.g. PREP with Dosino) and the exit track will be
started. The emergency stop is effective even if no user is logged in, e.g.
when the user has been logged out automatically.

The emergency stop has no effect on actions that may still be ongoing in
the manual control, each of which must therefore be stopped with
<Stop>.

Advertising