4.5 Determination overview

■■■■■■■■■■■■■■■■■■■■■■

320

■■■■■■■■

tiamo 2.4

Waiting and timeout periods
All waiting and timeout periods of commands are set to 0 during recal-
culating.

Non-executed commands
The commands REQUEST, WAIT, RECEIVE, SEND, TRANSFER, STIR,
SCAN, CTRL, MOVE, SWING, LIFT, PUMP, RACK, HEATER, FLOW, ELT,
ADD, LQH, PREP and EMPTY will be run through but not executed.

Report commands
The command REPORT is run through and a new original report is gen-
erated, but the defined output targets will be ignored. The command
REPORT only can be executed correctly during reprocessing when the
defined report template already has been used by a REPORT command
during the regular determination run. If the necessary report template
is not existing in the original determination, a message entry is created,
the REPORT command is skipped and the reprocessing will be contin-
ued.

Command variables
If a method contains calculations using command variables the deter-
mination can only be recalculated correctly if the DATABASE command
is executed after the corresponding commands.

Solution variables
If titrants are modified in the method then their associated variables
(e.g. Titer, Slope) will in any case (even when at the same time the
corresponding variables have been modified under Variables) be
taken from the (see Chapter 6.6.2.1, page 1374) and updated for the
recalculation.

Assign titer, common variables and global variables
If a determination without statistics is recalculated in which a titer, a
common variable or a global variable has been assigned, then the
assignment will only take place when the recalculation is confirmed
with [OK]. If a determination with statistics is recalculated in which a
titer, a common variable or a global variable has been assigned, then
the assignment with the data of the last recalculated determination will
only take place after the recalculation is confirmed with [OK].

4.5.2.6.4

Modifications

4.5.2.6.4.1

Modifications - Overview

Subwindow: Database

Determination overview Determinations Repro-

cess...

Reprocessing Modifications

In the subwindow Modifications of the dialog window Reprocessing
modifications can be made on the following 4 tabs:

Variables
Modification of the variables used in the selected determination.

Method
Modification of the method used in the selected determination.