Metrohm tiamo 2.4 Manual User Manual

Page 1005

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

989

Slope

Minimum difference between the slope before and after the break point.
The smaller the difference, the more break points will be found.

Input range

0.0 - 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer endpoints will be found.

Input range

2 - 20

Default value

5

Window

A range (window) can be defined on the measured value axis or on the
time axis. The break point evaluation will only be carried out in the
defined window. Only the first break point in the defined window will be
recognized.

Selection

Measured value | Time | off

Default value

off

Lower limit

Value for the lower limit of the window.

Time for the lower limit of the window.

Window = Measured value
Input range

-200.0 - 200.0 µA

Default value

-200.0 µA

Window = Time
Input range

0 - 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Value for the upper limit of the window.

Window = Measured value
Input range

-200.0 - 200.0 µA

Default value

200.0 µA

Window = Time
Input range

1 - 999999 s

Default value

999999 s

Advertising