5 Method

tiamo 2.4

■■■■■■■■

561

Gran evaluation

NOTE

This evaluation method is possible only with Titrando and 855.

Gran evaluation

on | off (Default value: off)

If this option is enabled, then titration curves will be evaluated in accor-
dance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 932).

Initial volume

Volume which is present in the measuring vessel before the command is
started.

Input range

0.01 - 9999.99 mL

Default value

50.00 mL

Lower limit

Value for the lower limit of the Gran evaluation range.

Input range

-200.0 - 200.0 µA

Default value

-200.0 µA

Upper limit

Value for the upper limit of the Gran evaluation range.

Input range

-200.0 - 200.0 µA

Default value

200.0 µA

5.6.3.2.5.8

DET Upol - Additional measured values

Tab: Method

DET Upol Properties... Additional measured values

Command name

Name of the command.

Entry

25 characters

A maximum of 6 additional measured values can be defined on this tab.
These values can then be saved together in additional measured value col-
umns with the measured values present in the default settings.