Metrohm tiamo 2.4 Manual User Manual

Page 728

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

712

■■■■■■■■

tiamo 2.4

Titrando, 855
Input range

1 - 999 µL/min

Default value

20 µL/min

Drift correction

If this parameter is set to automatic or manual the drift correction ( =
Drift value × Time for drift correction) caused by empty consumption
during titration is deducted from the total consumption. The Time for
drift correction
is the period between end of conditioning and end of
titration.

Selection

automatic | manual | off

Default value

off

automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Input range

0.0 - 99.9 µL/min

Default value

0.0 µL/min

Stop volume

Stops when the given volume has been added after the start of the titra-
tion. The stop volume should be adapted to the size of the titration cell in
order to prevent any overflow.

Titrando, 855
Input range

0.000000 - 9999.99 mL

Default value

20.00 mL

Selection

off

Default value

off

Stop time

Determination stops when the specified time has elapsed since the start of
the conditioning.

Titrando, 855
Input range

1 - 999999 s