Metrohm tiamo 2.4 Manual User Manual
Page 1437
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.4
■■■■■■■■
1421
Input range
–2.147E+9 - 2.147E+9
Default value
7.500
Intervention limits for pH(0) (for pH electrodes only)
on | off (Default value: off)
If this option is activated, then the values for pH(0) will be shown in red
if the limits are exceeded.
Lower limit
Lower intervention limit for pH(0).
Input range
–2.147E+9 - 2.147E+9
Default value
6.000
Upper limit
Upper intervention limit for pH(0).
Input range
–2.147E+9 - 2.147E+9
Default value
8.000
Warning limits for E(0) (for ISE electrodes only)
on | off (Default value: off)
If this option is activated, then the values for E(0) will be shown in
orange if the limits are exceeded.
Lower limit
Lower warning limit for E(0).
Input range
–2.147E+9 - 2.147E+9 mV
Default value
–15 mV
Upper limit
Upper warning limit for E(0).
Input range
–2.147E+9 - 2.147E+9 mV
Default value
15 mV
Intervention limits for E(0) (for ISE electrodes only)
on | off (Default value: off)
If this option is activated, then the values for E(0) will be shown in red if
the limits are exceeded.
Lower limit
Lower intervention limit for E(0).