9 det u - measured value window – Metrohm tiamo 2.4 Manual User Manual

Page 525

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

509

5.6.3.2.3.9

DET U - Measured value window

Dialog window: Method

DET U Properties... Potentiometric evalua-

tion

Evaluation with measured value window (U) [New/[Properties] Mea-

sured value window #

Measured value windows are regions (windows) on the measured value
axis for which different parameters for potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.

Lower limit

Lower limit for the measured value window.

Titrando, 855
Input range

-2000.0 - 2000.0 mV

Default value

-2000.0 mV

Titrino
Input range

-2000 - 2000 mV

Default value

-2000 mV

Upper limit

Upper limit for the measured value window.

Titrando, 855
Input range

-2000.0 - 2000.0 mV

Default value

2000.0 mV

Titrino
Input range

-2000 - 2000 mV

Default value

2000 mV

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.

Titrando, 855
Input range

0 - 200

Default value

5

EP recognition

Filter for the recognition of equivalence points: