5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

690

■■■■■■■■

tiamo 2.4

Window

A range (window) can be defined on the measured value axis, on the vol-
ume axis or on the time axis. The break point evaluation will only be car-
ried out in the defined window. Only the first break point in the defined
window will be recognized.

Selection

Measured value | Volume | Time | off

Default value

off

Lower limit

Value for the lower limit of the window.

Window = Measured value
Input range

-2000.0 - 2000.0 mS/cm

Default value

-2000.0 mS/cm

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Window = Time
Input range

0 - 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Window = Measured value
Input range

-2000.0 - 2000.0 mS/cm

Default value

2000 mS/cm

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

9999.99 mL

Window = Time
Input range

0 - 999999 s

Default value

999999 s

5.6.3.3.6.8

MET Cond - Additional measured values

Tab: Method

MET Cond Properties... Additional measured values

Command name

Name of the command.

Entry

25 characters