5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
878
■■■■■■■■
tiamo 2.4
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
Measuring point recording
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Input range
0.1 - 999,999 s
Default value
2.0 s
Titrino
Input range
1 - 999,999 s
Default value
2 s
Start time
The measuring points are not entered in the measuring point list until this
time has elapsed since the start of the titration.
Input range
0 - 999,999 s
Default value
0 s
Start measured value pH
The measuring points are not entered in the measuring point list until the
start measured value has been reached.
Titrando, 855
Input range
–20.000 - 20.000
Selection
off
Default value
off
Titrino
Input range
–20.00 - 20.00
Selection
off
Default value
off
Start rate
The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. This parameter is not checked until
10 s after the start.