5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1162

■■■■■■■■

tiamo 2.4

not defined
No tests will be carried out.

Tandem dosing

on | off (Default value: off)

If this option is enabled, then two dosing devices are used for dosing
without interruption, meaning that while the first dosing device is filling
the second one is dosing and vice versa. This parameter is displayed only
for Titrandos, 846, 856, 867, 814, 815, 855, 864 and 874.

Dosing device

Selection of the number of the second dosing device (exchange or dosing
unit), with which the solution is to be dosed when the first dosing device
is not available.

Titrando, 846, 856, 867

Selection

1 | 2 | 3 | 4

Default value

2

814, 815, 855, 864, 874

Selection

1 | 2 | 3

Default value

2

Solution

Entry of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method run to see whether the
correct solution has been attached to the selected dosing device and
whether the dosing device type is correct. With nonintelligent exchange
or dosing units, only the cylinder volume is checked.

Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Filling rate

Rate at which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during fill-
ing to the largest possible value.