5 Method

tiamo 2.4

■■■■■■■■

1191

Titrando, 855
Selection

Solution name | not defined

Default value

not defined

not defined
No tests will be carried out.

Filling rate

Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during dos-
ing to the largest possible value.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Default value

maximum mL/min

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Titrino
Selection

1 | 2 | diff.

Default value

1

Sensor

Selection of a sensor of type pH electrode, metal electrode, pH elec-
trode
or ISE electrode from the list of sensors defined in the sensor
table. The calibration data for the sensor will be adopted for pH and ISE
electrodes.

Selection

Sensor name | pH electrode | Metal electrode |
ISE electrode

Default value

Metal electrode

Temperature measurement

Type of temperature measurement.