5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
668
■■■■■■■■
tiamo 2.4
Window = Measured value
Input range
-200.0 - 200.0 µA
Default value
200.0 µA
Window = Volume
Input range
0.00000 - 9999.99 mL
Default value
9999.99 mL
Window = Time
Input range
0 - 999999 s
Default value
999999 s
Gran evaluation
NOTE
This evaluation method is possible only with Titrando and 855.
Gran evaluation
on | off (Default value: off)
If this option is enabled, then titration curves will be evaluated in accor-
dance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 932).
Initial volume
Volume which is present in the measuring vessel before the command is
started.
Input range
0.01 - 9999.99 mL
Default value
50.00 mL
Lower limit
Value for the lower limit of the Gran evaluation range.
Input range
-200.0 - 200.0 µA
Default value
-200.0 µA
Upper limit
Value for the upper limit of the Gran evaluation range.
Input range
-200.0 - 200.0 µA
Default value
200.0 µA