5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

974

■■■■■■■■

tiamo 2.4

The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other mea-
sured quantities will be interpolated with a fix measured quantity (Mea-
sured value
or Time) for the fixed endpoint from the measuring point
list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.4.5.6, page 978).

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.5.6, page 978).

[Delete]

Delete the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the mea-
suring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

1.0 - 2000.0 mV/s

Default value

25.0 mV/s