8 Manual control

tiamo 2.4

■■■■■■■■

1619

Measuring temperature

Temperature, which can be entered manually. If a temperature sensor is
connected, the temperature is measured continuously and the value
entered will be ignored. The value is used for the temperature correction
of the measurement.

Titrando, 855
Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Titrino
Input range

–170.0 - 500.0 °C

Default value

25.0 °C

I(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.

Titrando, 855
Input range

–125.0 - 125.0 µA (Increment: 0.5)

Default value

5.0 µA

Titrino
Input range

–127 - 127 µA (Increment: 1)

Default value

5 µA

U(pol)

The polarization voltage is the potential applied to the polarizable elec-
trode during an amperometric measurement.

Titrando, 855
Input range

–1250 - 1250 mV (Increment: 25)

Default value

400 mV

Titrino
Input range

–1270 - 1270 mV (Increment: 10)

Default value

400 mV

Reference temperature

The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).

712
Input range

–170.0 - 500.0 °C

Default value

20.0