6 Configuration

tiamo 2.4

■■■■■■■■

1387

function is carried out, the rate will automatically be reduced to the high-
est possible value.

Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

Maximum dosing / filling rate for the exchange unit, depending
on the cylinder volume:

Cylinder volume

Maximum rate

1 mL

3.0 mL/min

5 mL

15.0 mL/min

10 mL

30.0 mL/min

20 mL

60.0 mL/min

50 mL

150.0 mL/min

NOTE

Enter lower rates for solutions of higher viscosity.

Tubing parameters
Definition of the length and diameter of the tubing connected to the
exchange unit.

NOTE

Default values have already been entered for the tubing parameters;
these correspond to the dimensions of the standard tubing supplied. If
you do not make any modifications to the tubing connections then you
do not need to modify the tubing parameters. For information on
length and diameter of other tubing see Products/Accessories and
columns
at

http://products.metrohm.com

.

Dosing tip

Length

Length of the tubing connecting the stopcock to the dosing tip.

Input range

0.0 - 999.9 cm

Default value

40.0 cm