6 Configuration
tiamo 2.4
■■■■■■■■
1387
function is carried out, the rate will automatically be reduced to the high-
est possible value.
Input range
0.01 - 150.00 mL/min
Selection
maximum
Default value
maximum
Maximum dosing / filling rate for the exchange unit, depending
on the cylinder volume:
Cylinder volume
Maximum rate
1 mL
3.0 mL/min
5 mL
15.0 mL/min
10 mL
30.0 mL/min
20 mL
60.0 mL/min
50 mL
150.0 mL/min
NOTE
Enter lower rates for solutions of higher viscosity.
Tubing parameters
Definition of the length and diameter of the tubing connected to the
exchange unit.
NOTE
Default values have already been entered for the tubing parameters;
these correspond to the dimensions of the standard tubing supplied. If
you do not make any modifications to the tubing connections then you
do not need to modify the tubing parameters. For information on
length and diameter of other tubing see Products/Accessories and
columns at
.
Dosing tip
Length
Length of the tubing connecting the stopcock to the dosing tip.
Input range
0.0 - 999.9 cm
Default value
40.0 cm