Metrohm tiamo 2.4 Manual User Manual
Page 543
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
527
Titrando, 855
Input range
0.00000 - 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Input range
0.000 - 9999.99 mL
Default value
100.00 mL
Selection
off
Stop measured value
Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.
Titrando, 855
Input range
-2000.0 - 2000.0 mV
Selection
off
Default value
off
Titrino
Input range
-2000 - 2000 mV
Selection
off
Default value
off
Stop EP
The titration is stopped when the specified number of equivalence points
has been found.
Input range
1 - 9
Default value
9
Selection
off
Volume after EP
When the number of equivalence points defined under Stop EP has been
found, this volume will be added. In this way you can see the curve shape
after the equivalence point is reached.
Titrando, 855
Input range
0.01000 - 9999.99 mL
Selection
off
Default value
off
Stop time
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.