5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1082

■■■■■■■■

tiamo 2.4

command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Entry

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

NOTE

If an existing solution is selected as the solution to be added then the
Concentration and Unit of this solution will be adopted for calculat-
ing the standard addition. If not defined is selected, then Concentra-
tion
and Unit must be entered on the Standard addition tab.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando, 867
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Sensor

Selection of a sensor of type ISE electrode from the sensors configured
in the sensor table.

Selection

'Sensor name' | ISE electrode

Default value

ISE electrode

Temperature measurement

Type of temperature measurement.

Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.