Metrohm tiamo 2.4 Manual User Manual

Page 682

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

666

■■■■■■■■

tiamo 2.4

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the mea-
suring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.5 - 10.0 µA/mL

Default value

5.0 µA/mL

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the mea-
suring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.5 - 10.0 µA/mL

Default value

5.0 µA/mL

Break point evaluation

NOTE

This evaluation method is possible only with Titrando and 855.

Break-point evaluation

on | off (Default value: off)

A break-point evaluation is used to determine sharp changes of direction
in the titration curve.

EP criterion

Measure of the minimum sharpness of the break-point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value

Advertising