4 stat - track call, 5 tandem dosing – Metrohm tiamo 2.4 Manual User Manual

Page 879

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

863

5.6.3.8.4

STAT - Track call

Dialog window: Method

STAT/STAT pH/STAT U Properties... Monitor-

ing

[New]/[Properties] Track call #

Monitoring

Selection of the quantity to be monitored. When its limit values are viola-
ted a track is started.

Selection

Measured value | Dosing rate | Temperature |
Any

Default value

Measured value

Any
Means any of the three quantities.

Limit exceeding

Selection of the exceeded limit for which a track is to be started.

Selection

Lower limit | Upper limit | Any | OK

Default value

Any

Any
The track will be started if either the upper or the lower limit is viola-
ted.
OK
The track will be started when the monitored quantities are once again
within the limit values (including hysteresis).

Track name

Selection of the track that is to be started automatically.

Selection

'Track name'

NOTE

If a track is called for that is already running then it will be allowed to
finish and then restarted when it is free again.

5.6.3.8.5

Tandem dosing

Menu item: Method

STAT/ADD/DOS General/Hardware

If the Tandem dosing check box is enabled for the commands STAT,
ADD or DOS, then a second dosing device can be defined in order to
enable uninterrupted dosing. Dosing is carried out with a combination of
two dosing devices so that the second dosing device is dosing while the
first one is being filled and vice versa.

Advertising