2 minimum/maximum evaluation – Metrohm tiamo 2.4 Manual User Manual

Page 946

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

930

■■■■■■■■

tiamo 2.4

In non-aqueous solutions the half neutralization potential (HNP) is
frequently used instead of the pK value. The HNP is evaluated in the same
way as the pK value.

NOTE

If a start volume is to be added then it must be smaller than 1/2 V(EP1).

5.6.3.10.2

Minimum/Maximum evaluation

Parameters: Method

DET/MET/SET/KFT/KFC/MEAS Properties... Additional

evaluations

Minimum/Maximum evaluation

For the minimum or maximum measured value the associated volume,
time and temperature are interpolated from the measuring point list.
The evaluation begins as soon as the slope of the curve exceeds a particu-
lar threshold value.