5 meas opt - additional measured values – Metrohm tiamo 2.4 Manual User Manual

Page 1072

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1056

■■■■■■■■

tiamo 2.4

Window = Time
Input range

0 - 999,999 s

Default value

0 s

Upper limit

Measured value for the upper limit of the window.

Time for the upper limit of the window.

Window = Measured value
Input range

0 - 4,000 mAU

Default value

4,000 mAU

Window = Time
Input range

0 - 999,999 s

Default value

999,999 s

5.6.4.14.5

MEAS Opt - Additional measured values

Tab: Method

MEAS Opt Properties... Additional measured values

Command name

Name of the command.

Entry

25 characters

A maximum of 6 additional measured values can be defined on this tab.
These values can then be saved together in additional measured value col-
umns with the measured values present in the default settings.

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.

[New]

Opens the Calc. measured value # dialog window, in which the param-
eters for the additional measured value can be entered (see Chapter
5.6.3.11.1, page 933)
.

[Properties]

Opens the Calc. measured value # dialog window, in which the param-
eters for the additional measured value can be edited (see Chapter
5.6.3.11.1, page 933)
.