5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1122
■■■■■■■■
tiamo 2.4
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
5.6.5.8
ELT LOOP
5.6.5.8.1
ELT LOOP - Overview
Dialog window: Method
▶ ELT LOOP ▶ Properties... ▶ ELT LOOP - 'Command
name'
Loop command for the electrode test (test of pH electrodes), used
together with the ELT MEAS command.
Electrodes
This command can be carried out with electrodes of the pH electrode
type.
Appearance
The command has the following appearance:
Parameters
The parameters for the ELT LOOP command are configured in the fol-
lowing dialog window:
■
Command variables
The following command variables are generated by the ELT LOOP com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identi-
fica-
tion
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
.ENP
Electrode zero point of the sensor used for the command (in
mV for ISE sensors)