5 Method

tiamo 2.4

■■■■■■■■

473

5.6.3.2.2.5

DET pH - Stop conditions

Tab: Method

DET pH Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters

Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Titrando, 855
Input range

0.00000 - 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Input range

0.000 - 9999.99 mL

Default value

100.00 mL

Selection

off

Stop measured value pH

Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.

Titrando, 855
Input range

-20.000 - 20.000

Selection

off

Default value

off

Titrino
Input range

-20.00 - 20.00

Selection

off

Default value

off

Stop EP

The titration is stopped when the specified number of equivalence points
has been found.

Input range

1 - 9

Default value

9

Selection

off