5 Method

tiamo 2.4

■■■■■■■■

807

Parameters
The parameters for the KFT Upol command are set on the following
eight tabs:

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

Control parameters
Control parameters for the endpoint (measured value for the endpoint,
titration rate, user-defined parameters, stop criterion).

Titration parameters
Parameters for the run of the titration.

Stop conditions
Definition of conditions which cause the titration to stop.

Conditioning
Enable/disable conditioning. Definition of start drift, drift correction
and stop conditions.

Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.

Additional measured values
Definition of additional measured values of other measuring com-
mands, which are saved as additional columns in the measuring point
list.

Command variables
The following command variables are generated by the KFT Upol com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':

Identification

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CBY

Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated

.COK

Command status; 1 = Condition requirement ful-
filled
, 0 = Condition requirement not fulfilled

.CONC

Concentration of the solution used for the command