Metrohm tiamo 2.4 Manual User Manual

Page 1077

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1061

Sensor

Colorimetric sensor

Selection of a colorimetric sensor from the table of colorimetric sensors.
The calibration data for the colorimetric sensor will be adopted for the
determination.

Selection

'Name of the colorimetric sensor' | empty or
'first list entry'

Default value

empty or 'first list entry'

Temperature measurement

Type of temperature measurement.

Displayed only when Device type = 089 Photometer.

Selection

continuous | off

Default value

continuous

Stirrer
The Stirrer section is only displayed for Device type = 089 Photome-
ter
.

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate.

Input range

0 - 15

Default value

5

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.