5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

812

■■■■■■■■

tiamo 2.4

890
Selection

1

Default value

1

Sensor

Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement.

Titrando
Input range

–1,250 - 1,250 mV (Increment: 25)

Default value

400 mV

Titrino
Input range

–1,270 - 1,270 mV (Increment: 10)

Default value

400 mV

Electrode check

on | off (Default value: off)

If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

Titrando
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature on the Titration parameters tab will be used.