Metrohm tiamo 2.4 Manual User Manual
Page 877
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
861
Troubleshooting
Problem
Remedy
The control point is not main-
tained well. The measured value is
at times too high and at times too
low. The controller "oscillates".
■
Reduce the Max. rate and
possibly also the Min. rate.
■
Set the Dynamics higher.
■
Check if stirring is effective.
■
Arrange the electrode and
buret tip to an optimum.
■
Possibly use an exchange/dos-
ing unit with a smaller cylinder
volume (smaller volume incre-
ment per pulse).
The control point is not reached
for too long.
■
Set the Dynamics lower.
■
Increase the Min. rate.
■
Increase the Max. rate.
The control range is not reached
quickly enough.
Set the setpoint value with a SET
pretitration in such a way that the
control range is reached more
quickly.
Commands
Depending on the measured value one of the following two STAT com-
mands can be selected: