5 Method

tiamo 2.4

■■■■■■■■

861

Troubleshooting

Problem

Remedy

The control point is not main-
tained well. The measured value is
at times too high and at times too
low. The controller "oscillates".

Reduce the Max. rate and
possibly also the Min. rate.

Set the Dynamics higher.

Check if stirring is effective.

Arrange the electrode and
buret tip to an optimum.

Possibly use an exchange/dos-
ing unit with a smaller cylinder
volume (smaller volume incre-
ment per pulse).

The control point is not reached
for too long.

Set the Dynamics lower.

Increase the Min. rate.

Increase the Max. rate.

The control range is not reached
quickly enough.

Set the setpoint value with a SET
pretitration in such a way that the
control range is reached more
quickly.

Commands
Depending on the measured value one of the following two STAT com-
mands can be selected: