Metrohm tiamo 2.4 Manual User Manual

Page 775

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

759

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the Fixed endpoint evaluation # dialog window for entering a
new fixed endpoint (see Chapter 5.6.3.4.5.10, page 761).

[Properties]

Opens the Fixed endpoint evaluation # dialog window to edit the
fixed endpoint selected in the table (see Chapter 5.6.3.4.5.10, page
761)
.

[Delete]

Deletes the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the mea-
suring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

1.0 - 2000.0 mV/mL

Default value

25.0 mV/mL

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the mea-
suring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Advertising