5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1194

■■■■■■■■

tiamo 2.4

Titrino
Input range

0.01 - 150.00 mL/min

Default value

maximum mL/min

NOTE

The dosing rate should be reduced for viscous liquids.

Dosing time

Time during which dosing is to be carried out. This parameter is only dis-
played for Dosing criterion = Volume/Dosing time or Dosing rate/
Dosing time.

Input range

0 - 9999999 s

Default value

100 s

Pause

Waiting period before dosing is started.

Input range

0 - 999999 s

Default value

0 s

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously. This value is used for
temperature correction during monitoring with pH measurements..

Titrando, 855
Input range

-20.0 - 150.0 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Input range

0.1 - 999999 s

Default value

5.0 s