3 kft upol - start conditions, Start conditions – Metrohm tiamo 2.4 Manual User Manual

Page 829

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

813

off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.

Stirrer

Stirrer

Selection of the stirrer.

Titrando
Selection

1 | 2 | 3 | 4 | off

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando
Input range

-15 - 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is activated, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos.

Switch on/off automatically

on | off (Default value: on)

If this option is activated, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command. This parameter is displayed only for 751, 785, 784,
795 and 799.

5.6.3.5.3.3

KFT Upol - Start conditions

Tab: Method

KFT Upol Properties... Start conditions

Command name

Name of the command.

Entry

25 characters

The start conditions are processed in the sequence shown before the titra-
tion is started.