Metrohm tiamo 2.4 Manual User Manual

Page 1333

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.4

■■■■■■■■

1317

Password policies

NOTE

If the Password monitoring by Windows option is enabled, then
the parameters for password protection are deactivated. Compliance
with the directives pursuant to 21 CFR Part 11 must then be ensured by
the administrator in Windows.

Enforce unique password

Activating this check box ensures that a user can use a given password
only once.

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)

Enforce use of special characters

Activating this check box ensures that the password must contain at
least one special character (@, #, ~, etc.).

Selection

on | off

Default value

off

Minimum password length

Activating this check box ensures that the password must contain at
least the indicated number of characters.

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)

Input range

1 - 10 characters

Default value

6 characters

Passwords expire every

Activating this check box ensures that the user must enter a new pass-
word before the validity period expires. When a user logs in whose pass-
word will expire within the next 10 days, a corresponding message will
appear. If the validity period has expired, the user can only log in if he or
she changes the password.

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)

Advertising