Metrohm tiamo 2.4 Manual User Manual

Page 375

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Database

tiamo 2.4

■■■■■■■■

359

Measuring input:

Number of the measuring input to which the sensor is connected (+ desig-
nation iConnect for intelligent sensors).

Slope:

Electrode slope of the sensor used (in % for pH sensor or in mV for ISE).

E (0):

Electrode zero point of the sensor used (in mV only for ISE).

pH (0):

Electrode zero point of the sensor used (dimensionless for pH sensor).

c (blank):

Blank value of the ISE sensor used.

Cell constant:

Cell constant of the conductivity sensor used.

Calibration temperature:

Temperature during the calibration.

Calibration date:

Display of date and time of the last calibration.

Calibration method:

Display of the method the calibration has been carried out with.

User:

Display of the user logged in during the calibration or who entered the
calibration data manually.

Measuring input

Display of the measuring input the calibration has been carried out with.

a ... e

Display of the calibration data a ... e (for the Thermoprobe only).

Common variable 'Name'
Information about the common variable used.

Value:

Value and unit of the common variables after the last value assignment.