5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

888

■■■■■■■■

tiamo 2.4

[New]

Opens the Fixed endpoint evaluation # dialog window for entering a
new fixed endpoint.

[Properties]

Opens the Fixed endpoint evaluation # dialog window to edit the
fixed endpoint selected in the table (see Chapter 5.6.3.4.3.10, page 717).

[Delete]

Deletes the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the mea-
suring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.1 - 20.0 pH/mL

Default value

1.0 pH/mL

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the mea-
suring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.1 - 20.0 pH/mL

Default value

1.0 pH/mL