Operating the System

Overview of Sample Analysis

The 800 system indicates that it is ready to analyze samples when the following
conditions exist:

S Ready screen appears
S sample door is open
S Analyze key is lit

Do not attempt to analyze samples if the system is not ready.

The system aspirates all samples except expired gas samples. Aspiration reduces
exposure to biohazardous materials and increases precision:

S by sampling the correct volume for each analysis
S by ensuring precise sample heating before sample analysis
S by reducing instrument-to-instrument and operator-to-operator variability

Identifying the Sample Device

The system automatically identifies the sampling device. If the system cannot
identify the device, a message box appears prompting you to select one of the
following sample devices:

S 1 mL syringe
S >2 mL syringe
S capillary

Select the device type and press Enter. Then press Analyze to begin analysis.

Detecting Bubbles

If the system detects bubbles when it aspirates a sample, the Bubbles Detected in
Sample message box appears. If the sample is not at point A or beyond, position
the sample manually to continue with analysis.

Press . . .

To . . .

Sample in Place

begin analysis after you move the sample to the measurement
module.

Cancel

cancel analysis. The system starts a wash and returns to the Ready
screen.