Bayer HealthCare Rapidlab 800 User Manual

Page 577

Advertising
background image

Performance Characteristics

EĆ51

Table EĆ46.āă800 COĆox Module Recovery and Precision with Total Hemoglobin of

18-22 g/dL and Oxyhemoglobin > 80%

Analyte

Mode

n

Mean

Reference

Mean Difference

from Reference

WRSD

*

tHb

Capillary

72

21.04

0.21

0.21

Syringe

68

20.72

0.19

0.22

FO

2

Hb

Capillary

72

90.03

–0.73

0.24

Syringe

68

92.31

–1.21

0.64

FCOHb

Capillary

72

6.56

0.53

0.16

Syringe

68

4.48

0.95

0.72

FMetHb

Capillary

72

3.41

–0.08

0.21

Syringe

68

3.21

–0.01

0.20

FHHb

Capillary

72

0.00

0.27

0.11

Syringe

68

0.00

0.28

0.16

* WRSD = within-run standard deviation