Bayer HealthCare Rapidlab 800 User Manual

Page 186

Advertising
background image

3Ć18

Rapidlab 800 Operator's Manual

Figure 3Ć6.āăOpening the Measurement Module Door

Push up the door latches

and lift the door.

3.

Verify that the reference sensor fill solution is not below the fill line. Refer to

Filling the Reference Sensor

, page 3-69.

CAUTION:

Do not remove or return the sensors to the measurement module

without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.

4.

If necessary, fill the sensor as described in

Filling the Reference Sensor,

page 3-69.

5.

Verify that the pH, K

+

, Cl

, and Ca

++

sensors are almost full with only a

bubble at the top and that the Na

+

sensor is completely full.

If any sensor is not properly filled, replace the fill solution as described in

Filling the Measurement Sensors

, page 3-80.

6.

Close the measurement module door.

7.

Press

Continue

and allow the system to warm up.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

If . . .

Then . . .

you want to perform a
two-point calibration

press

Yes

.

you do not want to perform a
two-point calibration

a. Press

No

.

b. Press

Home

to return to the Ready screen.

!