Fluid detector test – Bayer HealthCare Rapidlab 800 User Manual

Page 352

Advertising
background image

4Ć66

800 Series Operator's Manual

Fluid Detector Test

Use this procedure to test the ability of the fluid detectors to detect the presence of
fluids and to detect whether the fluids are clear or opaque. You can also test fluid
detectors 1, 1A, and 2 with whole blood by inserting a sample device containing
whole blood into the sample port.

1.

Access the Fluid Detector Test from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

1 Fluidics System

and press

Enter

.

c. Select

5 Fluid Detector

and press

Enter

.

The Fluid Detector Test screen appears as shown in Figure 4-8.

Figure 4Ć8.ā Fluid Detector Test Screen

Perform the

Baseline/Clear test.

Check here for results.

2.

Test the fluid detectors with clear fluid and air:

a. Select

Baseline/Clear

and press

Enter

.

b. Press

Start Test

.

c. Check the screen for the message FD1, FD1A, FD2, FD3, FD4, FD5

acceptable to verify correct fluid detector function.

d. Press

Exit Test

.

3.

Perform the appropriate action.

If . . .

Then . . .

the fluid detector failed the Baseline test

continue with step 6.

you want to test whole blood

continue with step 4.

you want to return to the Menu screen

continue with step 6.

5

1

Menu Code

3