Bayer HealthCare Rapidlab 800 User Manual

Page 208

Advertising
background image

3Ć40

Rapidlab 800 Operator's Manual

Figure 3Ć23.āăReplacing the Sample and Waste Pump Tubing

Waste Pump

Roller Cage

Platen

Tubing

Tubing Cuff

Sample Pump

Position 5 Connectors

Position 3

Connectors

Position 4

Connectors

a. Perform the appropriate step.

If you are replacing . . .

Then disconnect the . . .

sample pump tubing

tubing from both connectors at position 4 on the
reagent manifold.

waste pump tubing

tubing from both connectors at position 5 on the
reagent manifold.

CO-ox pump tubing

sample and waste tubing with two-way
connectors from the CO-ox pump as shown in
Figure 3-24.

3.

Replace the reagent pump tubing as shown in Figure 3-23:

a. Select the replacement tubing that has the tubing collar.

b. Grasp the left tubing cuff and pull it away from the platen.

c. While holding the tubing, turn the roller cage clockwise and gently pull the

tubing away from the platen and the roller cage.

d. Discard the tubing according to your laboratory protocol.

e. Hold the new tubing assembly with the tubing in front.