Bayer HealthCare Rapidlab 800 User Manual

Page 131

Advertising
background image

2Ć33

Operating the System

Message

Description

COox Cover Open During
Meas

The cover on the CO-ox module was open while the
system was analyzing the sample.

COox Sample Chamber
Temp Error

The CO-ox module sample chamber temperature is

±

0.35

°

C of range and cannot accept sample analysis

requests.

COox Sample Temp Out
of Range

The CO-ox module sample chamber temperature is not in
range at the end of a measurement sequence.

Edited Report

The Patient Information screen was edited after initial
completion.

Excessive Bubbles in COox
Sample

The CO-ox module detects a non-continuous fluid in the
sample path.

D2 Excessive Drift: __

The sensor drift is beyond predefined limits during a
one-point or a two-point calibration.

Data Entry Incomplete
__ Not Sent

Appears when required data entry fields were not
completed and Auto Send is turned on.

Insufficient COox Sample

Appears when there is not enough sample to fill the
CO-ox module sample chamber and the analysis cannot
be performed.

Insufficient Sample

Appears when there is not enough sample to fill the
measurement module and you manually positioned the
sample for measurement.

Interfering Substance
Detected

Substances that interfere with analysis are present in the
sample. Only appears when 270 CO-ox is attached.

Interfering Substance: tHb

The CO-ox module detects the presence of substances in
the sample that may interfere with the measurement.

Measurement Module
Temp Error

The measurement module temperature is not in range and
the system cannot accept sample analysis requests.

Measurement Module
Temp Warning

The measurement module temperature is

±

0.15

°

C of

range and can accept sample analysis requests.

__ Not Sent

Appears when you press Do Not Send at the end of
analysis.